(1)光源系統(tǒng)常見的故障
(1) Common faults of light source system
(a)光線發(fā)暗。排除了是光柵遮擋的原因外,基本可以肯定是燈泡接近使用壽命,更換新燈泡則可。氙光燈泡的使用壽命一般設(shè)計(jì)在500小時(shí),如果光源系統(tǒng)有壽命計(jì)時(shí)裝置,這個(gè)問(wèn)題就很容易解決。
(a) The light is dim. Excluding the reason that the grating is blocked, it is basically certain that the bulb is close to its service life, and it is OK to replace it with a new bulb. The service life of xenon light bulb is generally designed to be 500 hours. If the light source system has a service life timing device, this problem can be easily solved.
(b)開機(jī)燈泡不亮。如果排除了燒掉燈泡這個(gè)原因的話,那么問(wèn)題常常就出在主機(jī)高壓點(diǎn)火電路部分,因?yàn)殚_機(jī)的時(shí)候機(jī)器會(huì)產(chǎn)生上萬(wàn)伏特的脈沖高壓,元件會(huì)老化得特別快。注意此時(shí)不能用萬(wàn)用表去檢查脈沖高壓。
(b) The bulb doesn't light up when the machine is started. If the reason of burning the bulb is eliminated, the problem often lies in the high-voltage ignition circuit of the main engine, because the machine will produce a pulse high voltage of tens of thousands of volts when the machine is started, and the components will age very fast. Note that you can't use a multimeter to check the pulse high voltage at this time.
(2)人工氣腹系統(tǒng)常見的故障
(2) Common faults of artificial pneumoperitoneum system
(a)壓力顯示不正確。這多是氣體壓力傳感器損壞,如果機(jī)器設(shè)計(jì)有一組多個(gè)壓差壓力傳感器同時(shí)工作的,一般一個(gè)老化或者污染無(wú)法工作的時(shí)候,我們要把這組傳感器全部更換,否則只換其中一個(gè),其他的同組的傳感器壽命也接近臨界了。
(a) The pressure display is incorrect. This is mostly due to the damage of gas pressure sensors. If the machine is designed to have a group of multiple differential pressure sensors working at the same time, generally when an aging or pollution cannot work, we should replace all the sensors in this group, otherwise only one of them will be replaced, and the service life of other sensors in the same group will be close to the critical point.
(b)流量不準(zhǔn)確。氣體流量傳感器損壞更換即可。有時(shí)候是通過(guò)壓差來(lái)計(jì)算流量的,這時(shí)候我們就要更換新的壓差傳感器。
(b) The flow is not accurate. If the gas flow sensor is damaged, it can be replaced. Sometimes the flow is calculated by differential pressure. At this time, we need to replace the differential pressure sensor.
(c)設(shè)備漏氣。很多情況是器械密封圈老化而引起的。但也有設(shè)備機(jī)械減壓閥漏氣和電磁閥損壞的原因,這個(gè)比較容易判斷出并對(duì)其進(jìn)行修復(fù)。
(c) Air leakage of equipment. Many cases are caused by the aging of instrument sealing ring. However, there are also reasons for air leakage of mechanical pressure reducing valve and damage of solenoid valve, which is easier to judge and repair.
(3)液體加壓系統(tǒng)常見的故障
(3) Common faults of liquid pressurization system
(a)蠕動(dòng)泵損壞。
(a) The peristaltic pump is damaged.
(b)液體壓力傳感器損壞。其原理基本與氣體壓力傳感器相同。
(b) The liquid pressure sensor is damaged. Its principle is basically the same as that of the gas pressure sensor.
(c)設(shè)備內(nèi)部電路損壞。
(c) The internal circuit of the equipment is damaged.
(4)電凝電切系統(tǒng)常見的故障
(4) Common faults of electric coagulation and electric cutting system
(a)設(shè)備配件損壞。如腳踏開關(guān),負(fù)極板導(dǎo)線,連接導(dǎo)線等常用的易損配件。
(a) Equipment accessories are damaged, such as foot switch, negative electrode plate, conductor, connecting conductor and other common vulnerable accessories.
(b)電切電凝,雙極故障。在設(shè)備內(nèi)部,這些部分都由不同的電路模塊組成。我們可以根據(jù)其表現(xiàn)的故障現(xiàn)象對(duì)相應(yīng)的模塊電路進(jìn)行維修或更換。
(b) Electric switch, electrocoagulation and bipolar fault. In the equipment, these parts are composed of different circuit modules. We can repair or replace the corresponding module circuits according to the fault phenomena.