在醫(yī)院各科室中,醫(yī)療用內(nèi)窺鏡的使用非常嚴(yán)格,包括它的零部件使用,下面
內(nèi)窺鏡設(shè)備維修公司為您講解下相關(guān)內(nèi)容。
In all departments of the hospital, the use of medical endoscope is very strict, including the use of its components. The endoscope equipment maintenance company will explain the relevant contents for you.
室內(nèi)儀器定位放置:
Indoor instrument positioning:
手術(shù)間的電源插座應(yīng)為屋頂懸掛式,避免因工作人員走動(dòng)碰脫電源插座而使電源線受干擾,保證手術(shù)的順利進(jìn)行,同時(shí)也避免由于電壓不穩(wěn)而造成儀器損壞。
The power socket in the operating room should be roof-mounted to avoid the interference of the power cord caused by the staff moving and touching the power socket, so as to ensure the smooth operation and avoid the damage of the instrument due to the unstable voltage.
監(jiān)視器:
Monitor:
一般監(jiān)視器在安裝時(shí)已經(jīng)都是有相關(guān)技術(shù)人員進(jìn)過(guò)測(cè)試并調(diào)整到了很好的顯示狀態(tài),所以不建議使用者自行調(diào)試。如果監(jiān)視器不顯示,請(qǐng)檢查插線板和電源線的連接,顯示器開(kāi)關(guān)是否處于打開(kāi)狀態(tài)、攝像系統(tǒng)是否處于打開(kāi)狀態(tài)、攝像系統(tǒng)的輸出是否正確的連接到監(jiān)視器的輸入端口。
Generally, the monitor has been tested and adjusted to a good display state by relevant technical personnel during installation, so it is not recommended that users debug by themselves. If the monitor does not display, check the connection of the patch panel and the power cord, whether the display switch is on, whether the camera system is on, and whether the output of the camera system is correctly connected to the input port of the monitor.
顏色失真后可將攝像系統(tǒng)的光學(xué)接口對(duì)準(zhǔn)顏色比較鮮明的物體或圖片,通過(guò)監(jiān)視器上的菜單進(jìn)行對(duì)比度、飽和度、亮度的調(diào)節(jié)。避免強(qiáng)磁物體靠近監(jiān)視器,定期作消磁保養(yǎng),當(dāng)磁化嚴(yán)重或消磁失效的時(shí)候,請(qǐng)聯(lián)系廠家技術(shù)人員。
After color distortion, the optical interface of the camera system can be aimed at objects or pictures with bright colors, and the contrast, saturation and brightness can be adjusted through the menu on the monitor. Avoid strong magnetic objects close to the monitor, and perform degaussing maintenance regularly. In case of serious magnetization or degaussing failure, please contact the manufacturer's technical personnel.
攝像系統(tǒng)和光源:
Camera system and light source:
雖說(shuō)很多醫(yī)療內(nèi)窺鏡聲明光學(xué)接口是防水的,但光學(xué)接口和連接線不建議消毒,可用防菌套套上,每次使用時(shí)消毒防菌套,這樣可以大大增加光學(xué)接口的使用壽命,如必須消毒請(qǐng)用低溫熏蒸的方法。
Although many medical endoscopes declare that the optical interface is waterproof, it is not recommended to disinfect the optical interface and the connecting line. It can be put on the germ-proof sleeve and sterilize the germ-proof sleeve every time it is used, which can greatly increase the service life of the optical interface. If it is necessary to disinfect, please use the method of low temperature fumigation.
插口不能強(qiáng)插拔,要對(duì)準(zhǔn)十二點(diǎn)的位置進(jìn)行直插、直拔輕拿輕放,定期作白平衡校準(zhǔn)。防止光學(xué)接口接觸液體,即使聲稱(chēng)是防水的設(shè)備,所以不要將調(diào)焦頭及連線浸泡消毒,如不甚接觸葡萄糖一類(lèi)的液體請(qǐng)手術(shù)后立即用干布擦干凈光學(xué)接口表面液體。防止卡口結(jié)晶運(yùn)動(dòng)不靈活,如果出現(xiàn)異常請(qǐng)不要勉強(qiáng)使用。
The socket cannot be forcefully inserted and pulled out. It should be directly inserted and pulled out at the position of 12 o'clock, handled gently, and regularly calibrated for white balance. Prevent the optical interface from contacting with liquid, even if it is claimed to be waterproof equipment, so do not immerse and disinfect the focusing head and wiring. If it does not contact with liquid such as glucose, please wipe the liquid on the surface of the optical interface with a dry cloth immediately after the operation. Prevent inflexible movement of bayonet crystallization. If there is any abnormality, please do not force to use it.
光纜不可用浸泡方式或高溫高壓方法消毒,如需消毒請(qǐng)用低溫熏蒸或淡酒精擦拭表面的方法進(jìn)行消毒。使用者要細(xì)心記錄光源燈泡的使用時(shí)間,臨近使用壽命時(shí)準(zhǔn)備備用燈泡。光源燈泡有使用壽命的限制,建議不使用時(shí)立即關(guān)閉光源。
The optical cable cannot be disinfected by immersion or high temperature and high pressure. If it needs to be disinfected, please use low temperature fumigation or light alcohol to wipe the surface. The user should carefully record the service time of the light source bulb and prepare the spare bulb when it is close to its service life. The light source bulb has a limited service life. It is recommended to turn off the light source immediately when it is not in use.
內(nèi)窺鏡對(duì)使用環(huán)境的要求為室內(nèi)環(huán)境以暗室環(huán)境為主,避免強(qiáng)光直射,手術(shù)開(kāi)始關(guān)閉日光燈,用無(wú)影燈照明,術(shù)前室內(nèi)的溫度保持在22℃~25℃、濕度50%左右、防酸、防堿、防蒸氣。避免強(qiáng)電、電磁場(chǎng)的干擾。更多事項(xiàng)就來(lái)我們網(wǎng)站
http://foxnewsss.com咨詢(xún)了解!
The requirements for the use environment of the endoscope are that the indoor environment is mainly darkroom environment, avoiding direct sunlight, turning off the fluorescent lamp at the beginning of the operation, using shadowless lamp for lighting, and maintaining the indoor temperature at 22 ℃~25 ℃, humidity at about 50%, acid, alkali and vapor proof. Avoid interference of strong current and electromagnetic field. Come to our website for more matters http://foxnewsss.com Consult and understand!